Contact an Account ManagerLiquid Chromatography Determination of the Anti-Androgen Vinclozolin and its Metabolites in Rat Serum
Summary
Type: HPLC
Application #: 120260
Catalog#: 720936.46 - MN HPLC Column, Nucleosil C18 AB, 100A, 5um, 4.6x250mm
Detection: UV, 212 nm
Conditions: Eluent:
A: methanol/acetonitrile (70:30) ,
B: 0.05 M NaH2PO4, pH=3.3 adjusted with 1 M HCl
Gradient: 40 - 30% B in 20 min, 30 - 25% B in 5 min
Flow rate: 1 ml/min
Authors: Sierra-Santoyo A Barton HA Hughes MF
Source: J Chromatogr B 809 (2004) 105-110